Tablet Instructions
31
RNS
®
System Programming Manual
Select a thumbnail to open it in a larger size, as shown below. ECoGs highlight detection and stimulation
events. Detected activity appears in blue. Undetected activity appears in black. Colored flags mark
specific points of detection or stimulation. You can
Z
OOM
in on the ECoG and increase the
G
AIN
. Use the
A
SSIGN
C
ATEGORY
field to assign the ECoG to a category. Select the
S
TAR
button at upper right to specify
this ECoG as a favorite, for fast access later using the
F
AVORITES
filter on the
EC
O
G
L
IBRARY
screen.
Figure 20 Viewing a User saved ECoG
Captured ECoGs reside in the RNS
®
Neurostimulator memory until they are transferred to the
programmer at the next interrogation with the wand, and are then cleared from memory. If
neurostimulator memory capacity is reached, captured ECoGs are overwritten by subsequent ECoGs.
The number of ECoGs that can be stored depends on the duration of each ECoG, which you can specify
in the
C
APTURE
W
INDOW
field in the
D
URATION
area on the far upper right side of the
EC
O
G
C
APTURE
Selecting ECoG Capture Settings
on page
L
IVE
EC
O
G
S
S
CREEN
Note: You can access the L
IVE
EC
O
G
S
screen only after interrogation of the neurostimulator, and
the wand must be held within approximately 1 inch of the neurostimulator to acquire live ECoGs.