Tablet Instructions
79
RNS
®
System Programming Manual
The
P
ROGRAMMING NEUROSTIMULATOR
dialog shows programming progress until complete.
Figure 65 Programming neurostimulator dialog shows progress of programming
Recommended Initial Responsive Therapy Settings
Responsive stimulation therapy should be enabled once the physician has determined the
electrographic activity of interest is being detected.
•
Program
B
URST
1
for electrodes on one lead and
B
URST
2
for electrodes on the other lead.
•
C
URRENT
= 1.0 mA.
•
Leave other parameters at their default values:
•
F
REQUENCY
= 200 Hz
•
PW
PER
P
HASE
= 160 µs
•
B
URST
D
URATION
= 100 ms
•
Apply the same settings to all five stimulation therapies: select the
Y
ES
button next to
A
LL
SETTINGS SAME AS
S
TIM
1
, near top center of the screen.
In general, stimulation should be delivered to the leads and electrodes from which electrographic
patterns of interest are observed. For example, if electrographic activity of interest is observed on all
channels, then the stimulation pathway should be configured to stimulate across all electrodes.
However, if electrographic activity of interest is observed on only 2 channels, then the stimulation
pathway should be configured such that current is delivered through only those electrodes with
electrographic activity of interest.