Tablet Instructions
46
RNS
®
System Programming Manual
2. On the
N
EUROSTIM
S
ETTINGS
screen, select
P
ATTERN
D
ETECTION
at upper center. A list of
three
C
OMPATIBLE
D
ETECTION
S
ETS
appears.
3. Select one of the three default detection settings. In the usual case, a patient has two active
leads to be configured, and therefore the recommended default pattern is
P
OWER CHANGE
75%
ON
C
H
1
&
3
(
DEFAULT
)
, the second item in the list. (If only one lead is implanted, this
two-channel option will not appear in the list.) The other default patterns detect on only one
lead.
•
Power change 75% is a detection setting that captures any large changes in the signal
that clinicians can program and optimize as detection patterns.
4. Select
L
OAD
S
ETTINGS
at lower right, then
R
EVIEW
&
P
ROGRAM
at upper right on the next
screen. The
R
EVIEW
screen opens.
5. On the
R
EVIEW
screen, make sure
D
ETECTION
is
E
NABLED
. Then, if you are satisfied with the
new settings, place the wand within approximately 1 inch of the neurostimulator and select
C
ONFIRM
P
ROGRAMMING
. If not satisfied, select
C
ANCEL
to return to the previous screen.