Tablet Instructions
75
RNS
®
System Programming Manual
The
A
DVANCED
S
ETTINGS
screen has upper, lower and bottom portions that let you select the
following:
Upper portion—Adaptive Therapy:
Adapts the frequency and the minimum amplitude of the
stimulation therapy as a percentage of the detected signal amplitude over a chosen channel, for
one or both bursts. The available options are:
Figure 60 Adaptive therapy settings—upper portion of Advanced Settings dialog
•
A
DAPT
C
HANNEL
:
Select one of four channels over which to enable adaptive therapy. By
default, this is
U
NUSED
, which means adaptive therapy is disabled.
•
A
DAPT
M
IN
A
MPLITUDE
:
Select a percentage of the detected signal amplitude to be used
as the minimum therapy amplitude.
Then for
B
URST
1
and/or
B
URST
2
, select settings for the following:
•
A
DAPTATION
:
Turn
O
N
or
O
FF
adaptive therapy for this burst.
•
T
HERAPY
F
REQUENCY
M
ULTIPLIER
:
A greater multiplier increases frequency of the burst.
•
M
AX
T
HERAPY
F
REQUENCY
:
Sets maximum frequency for this burst.
•
M
IN
T
HERAPY
F
REQUENCY
:
Sets minimum frequency for this burst.
•
D
EFAULT
T
HERAPY
F
REQUENCY
:
Sets default frequency for this burst.
Summary of Contents for PDMS 4340