![background image](http://html1.mh-extra.com/html/neuropace/pdms-4340/pdms-4340_user-programming-manual_1676339068.webp)
Tablet Instructions
64
RNS
®
System Programming Manual
S
HARE AND
/
OR
D
OWNLOAD A
D
ETECTION
S
ET
This section explains how to share a detection set with the users or programmers you select, and how to
download a shared detection set. For example, use this feature when more than one physician is treating
a patient.
You must be connected to the PDMS (via the tablet or a web browser on a personal computer) to define
or modify a custom detection set, and by default, these are saved to the PDMS only—not locally. If you
are connected to the PDMS during programming, you can program the neurostimulator with any detection
set saved under the current username or saved to the programmer you are using. However, if you are not
connected to the PDMS (offline) during programming, you can program only the pre-programmed default
detection sets or those detection sets previously downloaded from the PDMS to the programmer.
Steps to Share a Detection Set
1. Whenever connected to the PDMS, either on a programmer or using a web browser on a
personal computer, you can share the currently selected detection set by selecting the
S
HARE
button on the
N
EUROSTIM
S
ETTINGS
>
P
ATTERN
D
ETECTION
screen. The
S
HARE
D
ETECTION
S
ET
dialog opens.
Figure 45 The S
HARE
D
ETECTION
S
ET
dialog
Summary of Contents for PDMS 4340