Tablet Instructions
77
RNS
®
System Programming Manual
Figure 62 Post-episode interval settings—bottom portion of Advanced Settings dialog
Save Advanced Settings:
Select
D
ONE
to save the new advanced settings and return to the
S
TIMULATION
screen.
Test Stimulation Before Enabling
Before enabling responsive therapy, it is important to test stimulation to make sure it is well tolerated
by the patient and there are no undesired changes in the ECoG, such as afterdischarges.
Note: The system cannot test stimulation during a detected episode or during therapy delivery. To
test stimulation, either wait for the episode or therapy to complete, or disable detection and
therapy.
1. Select
T
EST
on the far left of the row to open the
T
EST
S
TIMULATION
screen and test the new
stimulation settings. Note that the
T
EST
button is not available if adaptive therapy is on (see
above). An ECoG is running when the screen opens. Select
D
ELIVER
to deliver the test burst. Select
S
TOP
/S
TORE
when desired to stop recording.
Figure 63 Test Stimulation screen
•
You can adjust settings and retest while this screen is open. To exit test stimulation, select
A
CCEPT
C
HANGES
or
C
ANCEL
C
HANGES
at upper right.
•
Stimulation may evoke visual, olfactory, auditory, taste, sensory or motor responses.