Tablet Instructions
62
RNS
®
System Programming Manual
Area Detection Tool
The
A
REA
detection tool can be selected only manually from the
D
EFINE
P
ATTERNS
dialog (select
M
ORE
C
ONTROLS
>
A
DV
C
ONTROLS
). It is not used as a suggested detector by the software. The
A
REA
detection tool measures the area between the ECoG and the time axis (see
).
Figure 43 Area detection tool example
The Area detector operates by comparing the normalized ECoG area (area per second) in a short-
term trend window (typically 1 to 4 seconds) to the normalized ECoG area in a long-term trend
window (typically 1 to 2 minutes). Detection occurs when the difference between the short-term and
long-term trend windows exceeds a programmed threshold (typically 37.5% to 100%).