Tablet Instructions
83
RNS
®
System Programming Manual
5. Select
R
EVIEW
&
P
ROGRAM
. The
R
EVIEW
screen opens, showing all the settings that will be
programmed.
•
If you want to enable detection and stimulation at this time, make sure
D
ETECTION
and
S
TIMULATION
are set to
E
NABLED
at upper center.
6. Place the wand within approximately 1 inch of the neurostimulator and select
C
ONFIRM
P
ROGRAMMING
. The
P
ROGRAMMING NEUROSTIMULATOR
dialog shows programming progress until
complete. The imported settings are deleted from the tablet when programming is complete. If
you do not complete this step, the settings will remain available on the tablet.
Note: After implanting a replacement neurostimulator, monitor the detection behavior. It may be
necessary to adjust the detection settings to ensure desired performance. If a bandpass detection
tool (R
HYTHMIC
A
CTIVITY
or S
PIKE
A
CTIVITY
) is enabled and the settings need to be adjusted, begin
by modifying the amplitude settings. See Recommended Therapy Modifications After
Observing Clinical Response on page 80.