![background image](http://html1.mh-extra.com/html/neuropace/pdms-4340/pdms-4340_user-programming-manual_1676339106.webp)
Specifications and Characteristics
102
RNS
®
System Programming Manual
Table 2. Guidance and manufacturer’s declaration –
electromagnetic immunity – for all equipment and systems
Immunity test
Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge
(ESD)
(IEC 61000-4-2)
±2, 4, 6, 8kV contact discharge
a
± 2, 4, 8, 15kV air discharge
a
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient /
burst
(IEC 61000-4-4)
± 2 kV for power supply lines
± 1 kV for input / output lines
b
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
(IEC 61000-4-5)
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
c
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
(IEC 61000-4-11)
0% U
T
0.5 cycle
0% U
T
1 cycle
d
40% U
T
5 cycle
d
70% U
T
25 cycles
0% U
T
5 Sec
If the user of the programmer and wand
requires continued operation during
power mains interruptions, it is
recommended that the programmer and
wand be powered from an uninterruptible
power supply or a battery.
Magnetic immunity (IEC
61000-4-8)
3, 30
e
A/m
a
Immunity to ESD tested to ±8 kV contact and ±15 kV air with Tablet (model 5000), Remote Monitor
(model 5100), and Wand (model W-02).
b
Immunity to electrical fast transients on I/O lines tested with Tablet (model 5000), Remote Monitor
(model 5100), and Wand (model W-02).
c
Immunity to surge line to earth tested with Programmer (model PGM-300) and Remote Monitor
(model DTR-300-E).
d
Immunity to voltage dips with compliance to 0% U
T
1 cycle for Tablet (model 5000), Remote Monitor
(model 5100), and Wand (model W-02) and to 40% U
T
5 cycle for Programmer (model PGM-300)
and Remote Monitor (model DTR-300-E).
e
Magnetic immunity tested with Tablet (model 5000), Remote Monitor (model 5100), and Wand
(model W-02) to 30 A/m.