![background image](http://html1.mh-extra.com/html/neuropace/pdms-4340/pdms-4340_user-programming-manual_1676339020.webp)
Tablet Instructions
16
RNS
®
System Programming Manual
I
NTERROGATE THE
RNS
®
N
EUROSTIMULATOR
To interrogate the RNS
®
Neurostimulator means to identify it, perform routine tests and measurements,
and retrieve currently programmed settings and stored data from it. Follow these steps to interrogate:
1. Connect the wand to the USB port on the tablet.
Figure 4 Tablet with connected wand
2. Place the wand within approximately 1 inch of the neurostimulator, concave side of wand
facing the neurostimulator. Then select
I
NTERROGATE
from the Start screen (left tile—see
). You can also select the green
I
NTERROGATE
tile at upper right in the
H
OME
screen
.) The wand must be held in place over the neurostimulator for interrogation to
succeed.
Figure 5 Interrogating neurostimulator dialog—interrogation in progress
Summary of Contents for PDMS 4340