![background image](http://html1.mh-extra.com/html/neuropace/pdms-4340/pdms-4340_user-programming-manual_1676339086.webp)
Tablet Instructions
82
RNS
®
System Programming Manual
Import Neurostimulator Settings
1. Interrogate the new neurostimulator, the one
to
which you will transfer settings.
2. Select the
U
TILITIES
(toolbox) icon at upper right. The
U
TILITIES
dialog opens (see figure above).
3. Select
I
MPORT
D
EVICE
. The
I
MPORT
D
EVICE
S
ETTINGS
dialog appears. It shows the device settings
available for import by device serial number and the date and time of last interrogation before the
settings were exported to the tablet. It also notifies you if the tablet has no settings that came
from a compatible neurostimulator (for example, if you exported from an RNS-320 for use on an
RNS-300M).
Figure 67 Import Device Settings dialog
Note: Device settings can be imported only once because they are deleted from the tablet once
you successfully import and select C
ONFIRM
P
ROGRAMMING
from the R
EVIEW
screen, as described
below.
4. Select the device settings—the selected settings will be highlighted—and then select
I
MPORT
. The
dialog notifies you that it is importing device settings.
When complete, the tablet Home screen opens. It will include a red
R
EVIEW
&
P
ROGRAM
tile at
lower right.
Figure 68 Home screen with the red Review & Program tile at lower right
Summary of Contents for PDMS 4340