Tablet Instructions
70
RNS
®
System Programming Manual
The following figure identifies the parameters of a burst of stimulation therapy.
Figure 55 Parameters of a single burst of stimulation
C
ONFIGURE
R
ESPONSIVE
T
HERAPY
Detection must be enabled before responsive therapy can be enabled. Before programming responsive
therapy, at least one detector must be defined in a detection set. (If you select therapy specific to a
detection pattern, at least one detector must be defined for each pattern.) To set up a therapy sequence,
determine the number of therapies desired for delivery in response to an episode. Select each therapy
and configure each burst, which is a collection of one to many biphasic pulses.
Note: For all settings, changed settings are indicated by blue buttons, with the new setting above
and the old setting below. Gray buttons indicate settings have not changed.
Note: It is recommended to review and perform detection analysis prior to enabling responsive
therapy.