USER MANUAL SWG
Page 93 of 129
Position 1: Acid-gas filter (#56795)
Required materials
2x Acid gas filter (#56795), contained in the service- set.
Required tools:
Needle-nose pliers.
Exchange interval:
Exchange necessary, if filter turns from purple to white.
Steps:
1. Remove exhausted acid gas filter from
the clips.
2. Pull the viton tubes from the
exhausted acid gas filter. If it is
necessary, use needle- nose pliers.
3. Plug the viton tubes on the new acid
gas filters. Be sure, that the tubing is
correct. (The single tubes are signed
-
side.)
Steps:
Pull the viton-tubes from the filter unit. If it is necessary, pliers can be used to solve
the tubes from the filter unit.
Remove the exhausted filter unit from the clip.
Plug the viton tubes on the filter unit. Push the filter unit on the clip.
- Socket
+ Socket
Remove
the
exhausted
filter
unit.
Install a new filter
unit.