USER MANUAL SWG
Page 85 of 129
Reference:
1. Name of the updateable device. In this example the multi gas bench (MGB).
2. FW update.
The chart below shows the different updateable systems.
Number
Device
1
MGB ( multi gas bench)
2
IO module
3
MCM (Measurement control modul)
4
Valve control module
5
Condensate pump module
6
Gas cooler module
7
ECM (electrochemical module)
8
Broadcast (virtual)
General steps for the firmware-update
1.
Copy the actual firmware for the analyser or the firmware from the option
on a SD-card. Be sure, that the firmware is in the mean root of the SD-card.
All
2.
Put the SD-card on the card slot on the operation unit. The card slot can
be found inside the door (see sketch below).
3.
If the SD-card is recognized, the analyser will make a noise.
4.
Open the path:
E
XTRA
/D
EVICE
I
NFO
. Dependent from the firmware update it
can be necessary to open the different submenus.
1
2
BEEP!!
Summary of Contents for SWG 100 SYNGAS
Page 1: ...SWG 100 SYNGAS USER MANUAL ...
Page 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Page 24: ...USER MANUAL SWG Page 24 of 129 ...
Page 32: ...USER MANUAL SWG ...
Page 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Page 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Page 99: ...USER MANUAL SWG Page 99 of 129 ...
Page 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Page 114: ...USER MANUAL SWG Page 114 of 129 ...
Page 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Page 127: ...USER MANUAL SWG Page 127 of 129 ...
Page 129: ...USER MANUAL SWG Page 129 of 129 ...