USER MANUAL SWG
Page 105 of 129
10.
Adjustment: NDIR-bench (MGB-Module)
ATTENTION
Close door
During the entire adjustment, the cabinet door must be
closed.
ATTENTION
The PIN-Code is: F1 F3 F2 F2 UP - DOWN
•
Select the menu:
E
XTRAS
/A
DJUSTMENT MENU
and input the 6 symbols
password.
•
gas selection (1)
choose between:
a.
Zero gas inlet.
Sucks gas from zero-gas inlet, if activated.
b.
Calibration gas inlet.
Sucks gas from calibration gas inlet, if
activated.
c.
Sample gas.
Sucks gas from sample-gas inlet, if
activated.
The screenshot below shows the different gas inlets and their position at
the analyser.
Zero-point: If this symbol appears, just wait until it disappears.
(2)
(2)
(1)