USER MANUAL SWG
Page 106 of 129
•
Mounting the adjustment gas cylinder: The pressure reducer must
be set at max. 500 mbar. The adjustment gas cylinder remains
closed.
•
Open NDIR cuvette menu:
Open the menu
E
XTRAS
/A
DJUSTMENT MENU
/N
DIR
ADJUSTMENT
. Here the values should now be close to zero. The menu
has the follow structure:
Reference:
1. Sample gas channel with current values (acutal value)
2. Lower set point, with set gas cylinder value and factor
3. Upper set point, with set gas cylinder value and factor
4. Manual execution of the zero point
5. Reset to standard adjustment.
Preferences: Enter the set point from the gas adjustment cylinder into the
menu (see 1). To do this, mark the lower set value and press the left and / or
right arrow key (see 2). A blue window appears in which the target value
can be entered (see 3). This must be done for both CH4 and CO2.