USER MANUAL SWG
Installation sample gas outlet (VENT) (see detail B-5)
•
WARNING
The gas outlet must not be closed.
•
WARNING
The tubing should be positioned so that the gas can escape from
the analyser almost without pressure.
•
WARNING
The line should therefore not be longer than 10 m and should be laid with
almost no kinks.
The sample gas flows into the sample gas inlet and out of the analyser again
after the analysis is completed. Because the sample gas can be toxic,
environmentally harmful or explosive, it is important to discharge this gas into a
secured environment.
Therefore, the following principle applies: Never let sample gas escape into the
installation room but lead it outside with a hose/pipe connection.