8.
Service and maintenance
0018
For a reliable function and high measurement quality it is necessary to inspect and
service the analyser regularly.
Besides the regular routine control by the operator (see chapter 8.1) the producer
recommends a regular half year maintenance, which must carried by a qualified
specialist.
8.1.
Preparing and information about the maintenance
It is important to power off the mains supply before the maintenance can be
started. Even if the main fuse is powered off, dangerous voltage is present.
It can be required to cut off the electric supply and safe this from an accidental
switch-on.
By maintenance works on the gas analyser dangerous and toxic gases may leak. The
gas supply must be cut-off.
It is important to comply with the national directives, which are the country
specific.
8.2.
Regular maintenance works by the operator
All inspections- and maintenance works are dependent from individual operating
conditions, and site. The specified intervals below are only benchmarks.
Review
Recommended
intervals
Actions
Moisture in the analyser.
Weekly
Remove the moisture. Call
vendor specialist.
Dirt and depositions in analyser.
Weekly
Remove the dirt, prevent
further penetration of dirt.
Inform vendor.
Dirt and moisture in the filter-
unit
Weekly
Exchange the filter- unit.
Testing the gas pipes of leakage
Weekly
Tighten or exchange the gas
pipe if it is necessary
Inspecting the conditions of the
gas filters
Every month
Exchange if necessary
In the follow chapters there will be introduced some service parts, which are
important for the reliable operation. These parts are independent from the regular
checks and must be replaced in a regular interval of minimum 6 month.
Summary of Contents for SWG 100 SYNGAS
Page 1: ...SWG 100 SYNGAS USER MANUAL ...
Page 12: ...USER MANUAL SWG Page 12 of 129 11 H3PO4 inlet G1 8 inner thread ...
Page 24: ...USER MANUAL SWG Page 24 of 129 ...
Page 32: ...USER MANUAL SWG ...
Page 63: ...USER MANUAL SWG Page 63 of 129 Detail A Alarm Relais ...
Page 89: ...USER MANUAL SWG Page 89 of 129 8 3 Description of the analyser 1 1 1 1 2 ...
Page 99: ...USER MANUAL SWG Page 99 of 129 ...
Page 110: ...USER MANUAL SWG Page 110 of 129 12 Appendix ...
Page 114: ...USER MANUAL SWG Page 114 of 129 ...
Page 116: ...USER MANUAL SWG Page 116 of 129 Positions of the components 3 4 ...
Page 127: ...USER MANUAL SWG Page 127 of 129 ...
Page 129: ...USER MANUAL SWG Page 129 of 129 ...