Leave the cycle phase detail and select the next phase. Select
with the
left/right arrow keys
the cycle phase detail. Adjust the phase-duration at 15 min. and
leave the cycle phase detail.
•
Switch to the next point and select with the
left/right arrow
key
the measurement SP1. Go in the cycle phase detail of the
measurement SP1. Here adjust the duration-phase 10 min. and
the pure measurement time: 5 min.
•
At last push
F3
key for insert a new phase and select with the
left/right
arrow
-
adjust the Purging time at 3 min. and the Quiet-time at 10 min.