937A Multi-Sensor System
A.19
DV-6M thermocouple sensor shielded cable with 9-pin "D" connector
Keyed "D" connector for
thermocouple
Pin
Description
1,5
thermocouple AC drive
2,4
thermocouple AC return
3
9
chassis ground
Cleaning the DV-6M Sensor
Roughing pump oils and other fluids condensing or decomposing on the
heated filament can contaminate the Sensor. This changes the emissivity
of the filament, which in turn can cause the calibration to change,
especially at low pressure.
It is not advisable to clean the Sensor. Trying to clean it may
either deform or break the filament. The deformed filament would
then cause additional error from a shift in the Sensor's output.
Replace the sensor if it becomes contaminated (see Spare Parts and
Accessories, page 67).