937A Multi-Sensor System
In either case, the control set point value is always displayed on the line for
the channel that is configured as the control sensor. When the selected
channel is not the same as the configured control channel, B1 or B2 will
appear on the selected display line of Slot
B to indicate which channel is the
configured control channel.
The display to the left shows an example of this type of configuration. The
control channel is B1. The
Gauge Select switch is set to channel
B2.Therefore, the value of the control setpoint is displayed on channel B1
and B1 is displayed on line B2 indicating that channel B1 is the control
channel.
3
Press the
Up or Down push-button to adjust the value accordingly.
Hold either button down to change it more rapidly.
A new set point takes effect and is saved in nonvolatile memory when either
the
Function Select or the Gauge Select switch is turned to another position.
The new set point is not saved if the Controller is turned off before turning
one of the switches.
To review a set point without affecting any output, turn the
Function
Select or Gauge Select switch to another position without changing
the displayed value.
Protection and Control Set Point Messages
Whether a cold cathode sensor or a control sensor is selected, bothcontrol
and protection set points will be
displayed. Only the set point
selected with the
Gauge Select
switch will be adjustable. If there is
no control sensor, only the
protection set point appears.
When a channel without a
protection or control set point is
selected, messages giving the
protection/control set point
configuration for each channel
appear.
Protection Set Point
Control Set Point
Protection and Control Set Point Display Messages
No
No on all display
CC
C1 on line B1 or B2
C2 on line B1 or B2
C12 on line B1 or B2
B1 on line B2 or
B2 on line B1
CONTROL 1 /
CONTROL 2
PRO
CTL
Channel does not have a protection or control set point.
When this message appears, select another channel to
display a set point
No cold cathode sensor module installed in Controller.
No protection or control set points are available.
Channel has a cold cathodesensor installed
Channel controls the cold cathode sensor in Slot CC.
Channel controls the cold cathode sensor in Slot A.
Channel controls the cold cathode sensor.
Channel B1 or B2 actually controls both cold cathode sensors,
regardless of Gauge Select switch and set point display position.
Control set point for slot CCI Slot A is active.
Selected set point is a protection set point.
Selected set point is a protection set point.