64
937A Multi-Sensor System
2
Press the
Up push-button to cancel this message.
The Controller will now work normally, except the correction factors are lost and
can no longer correct the affected channel's A/D readings. This can cause a
reading error up to +100%/-50% at pressures where sensor curves are relatively
flat (near the top or bottom of a pressure range for most sensors).
Setup Error
The message SETUP ERROR means at least one of two setup errors has
occurred.
An invalid capacitance manometer head DIP switch setting was
detected (see p. 29).
No recognized modules are plugged into the Controller.
Push the
Up push-button to cancel the message.
The Controller will ignore the incorrectly configured module. Note that this
does
not detect a mismatch between the DIP switch setting and the head
connected. This type of error cannot be detected, and an incorrect reading
will result.
It is possible that a capacitance manometer module with the first setup error
is plugged in or that a module of type other than cold cathode is in slot CC.
These are not recognized.
Pushing the
Up push-button will not cancel the message if no valid module is
in the Controller and pressure cannot be measured.