FrameMaker Ver.5.5E(PC) EP3000/EP3010 TROUBLESHOOTING
98.08.06
T-40
C06## Optical Malfunctions
✽
Code
Description
Detection Timing
C0600
Scanner Motor M4 mal-
function
•
When the Scanner is at a position other than home,
Scanner Reference Position Sensor PC81 does not
go from HIGH to LOW even after the lapse of 20
seconds after the Power Switch has been turned
ON.
•
When the Scanner is at the home position, PC81
does not go from LOW to HIGH even after the lapse
of 5 seconds after the Scanner has started a scan
motion.
•
PC81 does not go from LOW to HIGH even after the
lapse of 20 seconds after the Scanner has started a
scan motion.
C0610
Lens Motor M5 malfunction
•
The output from Lens Reference Position Sensor
PC90 does not go from HIGH to LOW, or vice
versa, even after the lapse of 15 seconds after M5
has started turning.
C0620
Mirror Motor M6 malfunc-
tion
•
The output from the Mirror Reference Position Sen-
sor PC86 does not go from HIGH to LOW even after
the lapse of 10 seconds after M6 has started turn-
ing.
•
The output from PC86 does not go from LOW to
HIGH even after the lapse of 3 seconds after M6
has started turning.
C06F0
SHOME signal failure
(Signal does not go LOW)
•
The SHOME signal does not go LOW even after the
lapse of 30 seconds after the SCAN signal has
gone LOW.
•
The BASE signal does not go LOW even after the
lapse of 30 seconds after the SCEND signal has
gone from LOW to HIGH.
•
In the F3 or F5 mode:
The SHOME signal does not go LOW even after the
lapse of 30 seconds after the ENABLE signal has
gone LOW.
C06F1
SHOME signal failure
(Signal does not go HIGH)
The SHOME signal does not go HIGH even after the
lapse of 7 seconds after it has gone LOW.
C06F2
BASE signal failure
(Signal does not go LOW)
The BASE signal does not go LOW even after the
lapse of 5 seconds after the ERASE OFF signal has
gone LOW.
C06F3
BASE signal failure
(Signal does not go HIGH)
The BASE signal does not go HIGH even after the
lapse of 5 seconds after the SCEND has gone HIGH.
C06F4
TRON signal failure
(Signal does not go LOW)
•
The TRON signal does not go LOW even after the
lapse of 5 seconds after the BASE signal has gone
LOW.
•
In the F3 or F5 mode:
The TRON signal does not go LOW even after the
lapse of 10 seconds after the SCAN signal has
gone LOW.
C06F5
TRON signal failure
(Signal does not go HIGH)
The TRON signal does not go HIGH even after the
lapse of 5 seconds after it has gone LOW.