Troubleshooting
6-3
6.2 Extending the Castors in Emergency
Figure 6-1 Extending the castors in emergency
1.
Open the seal cap at the base.
2.
Insert an 8mm Allen key into the opening.
3.
Turn the Allen key clockwise until the castors are extended (visual check). When the
castors are fully extended, it will be very difficult or impossible to turn the Allen key.
4.
Insert the seal cap.
5.
Move the defective operating table to the desired location and park it on its castors
(secure it, if required).
6.
Inform the customer service department.
In case the castors are extended do not press the locking button. Otherwise,
the castors cannot be extended.
Seal cap