5-1
5
Maintenance
5.1 Inspections and Maintenance
5.1.1 Inspections
Perform the following inspections each time before using the operating table, or at least once
a day.
1.
Has the operating table been properly cleaned and/or disinfected?
2.
Have the castors been properly locked?
3.
Are there any cracks on the pad?
4.
Have the pad been firmly fixed on the table top?
5.
Are there any damaged mechanical parts?
6.
Are there any damaged cables?
7.
Are all table-top movements functional?
5.1.2 Maintenance
1.
Complete electrical / safety inspections should be conducted every year by personnel
authorized by Nanjing Mindray.
2.
It is recommended that the operating table be serviced by professionals every year.
3.
It is recommended that the hydraulic oil be replaced by professionals every 5 years.
The operating table may be used only when properly lubricated and fully
functional. It is recommended that lubricate the operating table every year
by personnel authorized by Nanjing Mindray.
If the load on the operating table exceeds 230kg, half-yearly maintenance is
required.
Summary of Contents for HyBase 8300
Page 1: ...Operator s Manual HyBase 8300 HyBase 8500 Operating Table ...
Page 2: ......
Page 84: ...This page is intentionally blank ...
Page 106: ...This page is intentionally blank ...
Page 110: ...This page is intentionally blank ...
Page 122: ...This page is intentionally blank ...
Page 132: ...This page is intentionally blank ...
Page 142: ...This page is intentionally blank ...
Page 150: ...Appendices 10 8 3 Back plate up down up 90 5 down 45 5 ...
Page 158: ...Appendices 10 16 3 Back plate up down up 90 5 down 45 5 ...
Page 175: ...This page is intentionally blank ...
Page 176: ...P N 046 006753 00 2 ...