10-1
10
Appendices
A
Technical Specifications
A.1
Classification
A.1.1 According to Annex IX of the EU Directive 93/42/EEC
1.
Operating table: Class I
2.
Mobile Charging Station for IR Remote Control with LCD: Class II
A.1.2 According to the Type of Protection against Electrical Shock
Class I.
The operating table depends on the protective earth system to protect
against electrical shock. Before installing it, check and make sure that the
protective earth system in the operating room is reliable and safe.
A.1.3 According to Degree of Protection against Electrical Shock
1.
Applied part is classified as type B.
2.
Table pad is considered as applied part.
A.1.4 According to Mode of Operation
Intermittent: 10 minutes on /20 minutes off.
A.1.5 According to the Protection against Harmful Ingress of Water or
Particulate Matter
1.
Operating table: IPX4
2.
Foot switch: IPX8