Daily Operations
2-7
Collision protection is not active if the override panel is used. Avoid the
combination of extreme tilt and slope adjustments. Otherwise, it can cause
property damage.
If the batteries in the mobile operating table are exhausted, then the override
panel will only operate when the table is connected to the mains.
The override panel is intended for emergency use only.
The override panel employs the two-key-control mode. To use the panel,
press both the “Power On” key and the desired functional key
simultaneously.
2.3.2.2 Indicators
There are five indicators on the override panel: AC indicator, battery indicator, power indicator,
locking indicator and unlocking indicator, which have the same functions as those on the
corded hand control. See chapter 2.3.3.2 for details.