Arm Board
9-6
Be sure to check the arm board each time before use. Do not use worn or
damaged arm boards.
9.3.3.2 Installation and Use
1.
Hold the arm board with one hand.
2.
Hold the clamp with the other hand.
3.
Slide the clamp onto the side rail of the operating table and release the clamp.
4.
Pull the adjustment lever and move the arm board to the desired position.
5.
Secure the patient’s arm with the fastening strap.
6.
Make sure the arm board is reliably installed.
Make sure the clamp is tightly locked on the side rail and positioned upright.
Otherwise, the arm board can fall and cause personal injury and/or property
damage.
The maximum load of the arm board is 8.5 kg. Do not use it beyond this
limit.
Do not lean on the arm board during the surgery. Otherwise, the patient’s
arm can be permanently damaged.
Be sure to regularly check the position of the patient’s arm during the
surgery.
It is recommended the arm board be positioned parallel to the table top.
Otherwise, collision hazard is possible.
When using the arm board, be sure to cover it with a surgical drape.
Summary of Contents for HyBase 8300
Page 1: ...Operator s Manual HyBase 8300 HyBase 8500 Operating Table ...
Page 2: ......
Page 84: ...This page is intentionally blank ...
Page 106: ...This page is intentionally blank ...
Page 110: ...This page is intentionally blank ...
Page 122: ...This page is intentionally blank ...
Page 132: ...This page is intentionally blank ...
Page 142: ...This page is intentionally blank ...
Page 150: ...Appendices 10 8 3 Back plate up down up 90 5 down 45 5 ...
Page 158: ...Appendices 10 16 3 Back plate up down up 90 5 down 45 5 ...
Page 175: ...This page is intentionally blank ...
Page 176: ...P N 046 006753 00 2 ...