Daily Operations
2-34
14. Press and hold
to longitudinally slide the table top toward the foot end.
15. Press and hold
to longitudinally slide the table top toward the head end.
16. Press
or
to obtain the normal position or reverse position.
After adjusting the operating table, be sure to switch off the hand control.
Each time before using the operating table, check all functions by pressing
every functional key for 2 to 3 seconds. Make sure all the keys can work
normally. Do not use the damaged operating table.
If the operating table does not respond to the command even though the
function key is completely depressed, either re-aim the hand control or
move to a slightly different transmission position. In such cases the infrared
receiver at the column is being blocked by a person’s body.
When the IR remote control with LCD has been idle for about 90 seconds, it
will be switched off automatically.
If the table top is sloped (trendelenburg position) by more than 10
º
, it cannot
slide toward the head end.
10 minutes after the first low battery alarm, the operating table will power off
automatically.
It takes about 10 hours to fully charge the battery.
It is recommended that the IR remote control with LCD be slid into the
mobile charging station if it is idle.
Summary of Contents for HyBase 8300
Page 1: ...Operator s Manual HyBase 8300 HyBase 8500 Operating Table ...
Page 2: ......
Page 84: ...This page is intentionally blank ...
Page 106: ...This page is intentionally blank ...
Page 110: ...This page is intentionally blank ...
Page 122: ...This page is intentionally blank ...
Page 132: ...This page is intentionally blank ...
Page 142: ...This page is intentionally blank ...
Page 150: ...Appendices 10 8 3 Back plate up down up 90 5 down 45 5 ...
Page 158: ...Appendices 10 16 3 Back plate up down up 90 5 down 45 5 ...
Page 175: ...This page is intentionally blank ...
Page 176: ...P N 046 006753 00 2 ...