Arm Board
9-8
Do not use the cleaning agents or disinfectants containing alcohol where
such medical devices as electrosurgical units, defibrillators or defibrillator
monitors can be used. Otherwise, fire or explosion hazards are possible.
Do not use the disinfectants containing alcohol.
Do not use the disinfectants containing chlorine or compounds which can
release chlorine to disinfect the metal parts. Long-term use of these
disinfectants may corrode the metal parts, hence compromising safety and
effectiveness of the device. It is recommended that metal parts be
disinfected by 2% alkaline glutaraldehyde.
Be sure to only use the recommended disinfectants and methods.
Do not use scouring agents.
Do not use water containing metal particles, any metal tools (such as wire
brushes or steel wool), or products containing hydrochloric acid.
Do not disinfect the arm board mechanically.
9.4.3 Cleaning/Disinfecting Pad
Liquid can penetrate damaged pad and cause hygienic problems.
When disinfecting the patient’s skin, make sure the disinfectant does not
accumulate under the patient.
Do not use the worn or damaged pad.
Do not clean/disinfect the pad mechanically.
Do not use the disinfectants containing alcohol.
Do not use the disinfectants containing chlorine or compounds which can
release chlorine.
Summary of Contents for HyBase 8300
Page 1: ...Operator s Manual HyBase 8300 HyBase 8500 Operating Table ...
Page 2: ......
Page 84: ...This page is intentionally blank ...
Page 106: ...This page is intentionally blank ...
Page 110: ...This page is intentionally blank ...
Page 122: ...This page is intentionally blank ...
Page 132: ...This page is intentionally blank ...
Page 142: ...This page is intentionally blank ...
Page 150: ...Appendices 10 8 3 Back plate up down up 90 5 down 45 5 ...
Page 158: ...Appendices 10 16 3 Back plate up down up 90 5 down 45 5 ...
Page 175: ...This page is intentionally blank ...
Page 176: ...P N 046 006753 00 2 ...