A-5
other factors. Therefore, infusion sets of
different brand and model may differ in
maximum rates.
Note 2: The above declared rates are
based on BOON A2 and B.Braun Intrafix
Safeset infusion sets, and specified
infusion sets,. temperature of 20±2
℃
.
Bolus Rate/
Purge Rate
0.1-2000ml/h
/
VTBI
0.1-9999.99ml
---
Time
00:00:01-99:59:59 h:m:s
---
Volume
0.01-9999.99ml
0ml
Select Mode
Rate Mode, Body Weight Mode, Time
Mode, Sequential Mode, Loading Dose
Mode, Micro-infusion Mode and Ramp
Up/Down Mode
Rate Mode
Pressure
12 Levels are adjustable, respectively are
(75, 150, 225, 300, 375, 450, 525, 600,
750, 825, 900 and 975)mmHg
For 75mmHg, the error is
±50mmHg(6.7kPa); for other pressures,
the error is ±20% or ±113mmHg (15.1kPa),
whichever is larger.
Maximum occlusion pressure is about
1170mmHg.
Note 1: The detected pressure of the
infusion system (infusion pump and
infusion set) is affected by inner and outer
diameter, material, elasticity of the infusion
set and other factors. Therefore, infusion
sets of different brand and model may
differ in the detected pressure range.
Note 2: The above declared pressure is
based on BOON A2 and B.Braun Intrafix
Safeset infusion sets, and specified
infusion sets,temperature of 20±2
℃
.
450mmHg
Air detection
6 Levels are adjustable, respectively are
(20, 50, 100, 250, 500, 800) μl
Sensitivity of single air bubble is 20
μl.
Note 1: The detected air detection of the
infusion system (infusion pump and
infusion set) is affected by inner and outer
diameter, material, elasticity of the infusion
100
μl
Summary of Contents for BeneFusion VP5
Page 1: ...BeneFusion VP5 BeneFusion VP5 Ex BeneFusion VP5 Neo Infusion Pump Operator s Manual ...
Page 2: ......
Page 83: ...A 12 Unit of Rate ml h ...
Page 102: ...F 1 F Declaration of Conformity P N 046 013877 00 1 0 ...
Page 103: ......
Page 104: ......