A-15
A.6.3 Infusion Pump with PVC Specified Infusion Sets
Plotted on the basis of data collected over a two-hour measurement period.
Infusion set brand and model: SK ZPQ
Sampling quantity of pump: 3
Sampling quantity of infusion set: 3
Sampling interval:
△
t =0.5min
Test period: t =8h (1ml/h), 24h (25ml/h)
Infusion rate: Q (m/h)
Flow rate deviation over time (p
△
t)
Sampling interval:
△
t =0.5min
Observation windows: p
△
t = 2, 5, 11, 19, 31 mins
Maximum deviation over the course of a full observation window: Ep(Max) (%)
Minimum deviation over the course of a full observation window: Ep(Min) (%)
Average deviation: A (%) and B (%)