4-6
2. Adjust the slide clamp, insert
it into the operating lever; load
the set, confirm it is firmly
loaded into the tube slot, and
adjust or press the set at the
gray zone, to confirm the set is
loaded correctly.
3. Push the door latch and close
the door gently, the system will
enter the [
Set Selection
]
interface, indicating that the
infusion set is loaded correctly;
otherwise, it needs to be
reloaded.
WARNING
The infusion set should be firmly loaded into the slot, and not jutting on the
outside of the slot.
Before using this infusion pump, the infusion pump, infusion set and other
accessories should be loaded correctly.
Infusion pump with PVC specified infusion sets has to be used with PVC
specified infusion sets.
4.2.3 Change the Infusion Set
4.2.3.1 Infusion Pump with Conventional Infusion Sets
Follow the steps below to change the infusion set:
1. To prevent patient injury due to free flow, before changing the infusion set or
extruded tube, please shut down the Robert clip (or flow rate regulator). During
infusion, press
to stop the pump.
2. Pull the door latch up to open the door, pull the liquid check holder button upward,
and take out the loaded infusion set.