2-1
2
Overview
2.1 Description
2.1.1 Intended Use
The infusion pump is used in conjunction with the infusion set to control the dose of
liquid infused into the patient’s body. The pump may also be used for blood
transfusion. For this therapy only use disposables dedicated and labelled for
transfusion.
The infusion pump is suitable for adults, children and newborns in clinical
departments.
The infusion pump is expected to be used in institutes or units with healthcare
capabilities. This includes but is not limited to: outpatient departments, emergency
departments, wards, ICU, operating rooms, observation rooms, clinics, and nursing
hospital.
WARNING
The infusion pump is for clinical use. It must only be used under
appropriate conditions by professional clinicians, medical device
technicians, or by suitably trained nurses. Personnel using this product
must receive sufficient training. This product must not be operated by
anyone who has not been authorized to do so or has not received suitable
training.
2.1.2 Contraindications
None
2.1.3 Appearance, Parts and Features
The infusion pump primarily consists of a housing, a motor drive system, an input
system, a memory system, a control system, a display system, a sensor-based
monitor system and an alarm system. Eccentric camshaft is driven by stepper motor
during operation, making the fixed upper slider moves up and down sequentially and
regularly, infusion set is extruded regularly, and liquids in infusion set can flow
directionally at a certain rate, and all components are suitable for use in patient
environment. The wireless module and drop sensor are optional. Optional functions
Summary of Contents for BeneFusion VP5
Page 1: ...BeneFusion VP5 BeneFusion VP5 Ex BeneFusion VP5 Neo Infusion Pump Operator s Manual ...
Page 2: ......
Page 83: ...A 12 Unit of Rate ml h ...
Page 102: ...F 1 F Declaration of Conformity P N 046 013877 00 1 0 ...
Page 103: ......
Page 104: ......