7-2
Normally Open: Select when the hospital call system is set as [
NORM. Open
].
Alarm level: Three options: [
High
], [
Medium
] and [
Low
].The system sends nurse
call signals according to the alarm at the selected alarm level or above.
WARNING
Non-medical personnel are forbidden to modify the nurse call setting.
The nurse call function must be used in conjunction with a special cable.
NOTE
Medical personnel should not consider the nurse call function as the
main alarm notice approach, and rather combine the sound and visual
alarms of the infusion pump and the clinical performances and
symptoms of the patient in order to judge the patient's conditions and
take further attention as needed.
7.4 Wireless Networking (Optional)
The pump can be net-connected through built-in Wi-Fi module.
1. Select
[
Main Menu
]→[
User Maintenance
]→Input User Maintenance
Password→[
WLAN Setting
], then select [
On
] to enable Wi-Fi function.
2. Select [Ap Settings]:
Select [
SSID
]: Input the name of the to-be-connected network.
Select [
Password
]: If password is not required for to-be-connected network, you
can connect the network directly; If password is required for to-be-connected
network, please enter the password.
Select [
Encryption mode
]: Set the encryption mode of the to-be-connected
network, which shall be the same encryption mode as the Ap.
After the above settings are completed, select [
Confirm
] and press
.
3. Select [
Advanced Setting
], there are two ways to distribute IP address:
DHCP: IP address, Mask and Gateway can't be modified, automatically obtain an
IP address.
Manually: Enter IP address, Mask and Gateway.
Note: The wireless icon on the upper-right corner indicates the pump is configured
with wireless module and connected successfully. No icon indicates no wireless
module is configured or no connection.
After the pump and BeneFusion CS5 Infusion Supervision System (hereinafter called
CIMS) are communicated successfully over wireless network, the pump sends
real-time Bed No., version information, infusion parameters and other information to
CIMS, CIMS and the pump can display synchronously. For detailed descriptions,