4-10
Press
to continue the infusion.
4.2.12 BOLUS
In any run screen in the infusion mode, press
to enter the [
Bolus
] settings
screen. There are two ways to start the bolus:
Manual Bolus: Set [
Bolus Rate
], press and hold
to manual bolus, and
release to return to the original rate.
Auto Bolus: Set [
Bolus Rate
], [
Bolus VTBI
] and [
Bolus Time
], press
to
auto bolus.
NOTE
Patient's clinical condition and working condition of the infusion pump
must be monitored carefully.
4.2.13 Change the Rate during Operation
Rate may be modified without stopping the infusion. In any running screen of the
infusion mode, press
to change the value of the [
Rate
] into the adjustable state,
thus to set the expected rate, press
or
again for confirmation, then start to
infuse under the new set rate.
4.2.14 Complete
When [
VTBI
] is not set during the infusion and infusion is completed, if drop sensor is
installed and the switch of [
Drop sensor
] is on, the [
Empty
] alarm will be triggered.
When [
VTBI
] is set during the infusion and the remaining infusion time is close to the
[
Time Near End
] set by the users, the [
Time Near End
] alarm will be triggered. If no
action has been taken, the alarm will not be cancelled automatically until the infusion
is completed, and then switch to [
VTBI Complete
] or [
Empty
] alarm. Set [
Time Near
End
], please refer to
6.9 Time Near End
. When infusion is completed, enter to [
KVO
]
mode, and KVO mode will run for 30 mins at most. Infusion will stop automatically
after the KVO is finished, and the [
KVO Finish
] alarm will be triggered. Set KVO rate,
please refer to
6.1 KVO
.