4. Push the door latch and close the
door gently, the system will enter the
[
Set Selection
] interface, indicating
that the infusion set is loaded
correctly; otherwise, it needs to be
reloaded.
WARNING
The infusion set should be firmly loaded into the slot, and not jutting on the
outside of the slot.
Before using this infusion pump, the infusion pump, infusion set and other
accessories should be loaded correctly.
Infusion pump with silicone specified infusion sets has to be used with
silicone specified infusion sets.
4.2.2.3 Infusion Pump with PVC Specified Infusion Sets
System will inspect whether the set is loaded after completing the self-test:
If infusion set is not loaded, enter the infusion set [
Loading Guide
] interface; If
the infusion set is not required to load, press
to skip the step.
If conventional infusion set is loaded correctly or PVC specified infusion set is
loaded incorrectly, and door is closed, enter the [
Loading Error
] or [
Loading
Guide
] interface.
If PVC specified infusion set is correctly loaded and door is closed, and [
User
Maintenance
]→[
Brand Selection
]→[
On
], enter the [
Set Selection
] interface; if
the switch of [
Brand Selection
] is [
Off
], then skip the [
Set Selection
] interface.
Load infusion set according to the following method:
1. Pull the door latch up to open
the door.