A-6
set and other factors. Therefore, infusion
sets of different brand and model may
differ in the detected air detection.
Note 2: The above declared air detection is
based on BOON A2 and B.Braun Intrafix
Safeset infusion sets, temperature of
20±2
℃
.
Accumulate air
0.1-4ml
1.5ml
KVO Rate
0.1-5.0ml/h
0.5ml/h
Time Near End
Off, 1-30 min
when the time is <10min, step for 1min,
and step for 5min when the time is ≥10min.
3min
Max. Rate Limit 0.1-2000ml/h
2000ml/h
Drip Setting
Drip:10-60drop/ml
20drop/ml
gtt/min: On, Off
Off
Drip Abnormal
On, Off
Off
Empty
Sensitivity
High, Low
Low
Auto-lock Time
Off, 1-5min, step for 1min
Off
Reminder Time
Off, 1-5min, step for 1min
2min
Weight Unit
kg, lb
kg
Sound Volume
1-8
4
Brightness
1-8
4
Night Mode
On, Off
Off
Nurse call
On, Off
Off
History Record
Can store up to 2000 records.
/
Brand Selection On, Off
On
BW Mode
Configuration
Conc., Drug Amount and Volume
Conc.
Drug Library
On, Off
Off
Para. Memory
On, Off
Off
Date and Time
Time:_ _:_ _
00:00
Date:_ _ _ _-_ _-_ _
01-01-2018
Time format: 12h, 24h
24h
Summary of Contents for BeneFusion VP5