A-8
A.5 A Reference Table Showing Occlusion Alarm
Delay and Possible Dose
A.5.1 Infusion Pump with Conventional Infusion Sets
Occlusion Pressure
(mmHg)
Rate
(ml/h)
Time of Occlusion Alarm
(hh:mm:ss)
75
0.1
02:54:30
1
00:06:07
25
00:00:30
975
0.1
19:01:01
1
01:37:08
25
00:02:00
Pressure Setting
Rate
(ml/h)
Bolus Volume (ml)
75
25
<0.1
975
25
<0.15
NOTE
Test conditions:
FLUKE IDA4 PLUS tester
Infusion set brand and model: B.Braun Intrafix Safeset
Test temperature: 20±2
℃
Extension tube length: 1 meter
Occlusion alarm pressure, alarm delays and bolus volume may vary
depending on test conditions, temperature and tube length.
The above data are only typical values under normal test conditions. The
actual data may vary as test conditions change. Please refer to the test
data for the product you have purchased. Under the same standard
occlusion value and rate, the higher the value of the tested pressure is, the
longer the alarm time will be delayed.