3-3
3.1.3 Mount the Clamp
1. Align the mounting holes on the
installation dead plate with the
installation screw holes on the bottom
of the machine, and tighten the screws.
2. Turn counterclockwise to loosen the
pole clamp until an IV stand can be
inserted in.
3. Tighten the pole clamp clockwise to
firmly fix the device on the IV stand
(round vertical bar with diameter size of
15-32mm).
3.1.4 Multi-channel Pumps Combination
A maximum of 2 pumps can be combined (infusion pump and syringe pump can be
cross combined).
When multi-channel pumps are combined, align the rail on the bottom of the upper
pump to the rail slot on the top of the lower pump, slide from backward to frontward
until the alignment is matched, then strain the two pumps by the hasps. When
disconnecting, hold the upper-part of the hasp of the upper pump and pull out the
lower pump.