C-5
Alarm
Information
Alarm
Level
Reason
Countermeasure
[KVO Running]
Low
The alarm is triggered after
[VTBI Complete] alarm is
triggered and in KVO mode.
1. Press
to stop
infusion and cancel alarm.
2. Or the [KVO Finish]
alarm is triggered when
reach the KVO infusion
time.
[Infusion
interrupted]
Low
The alarm is triggered after
any of the [Occlusion], [Air in
line], [Accumulate air], [Door
opened], [No Battery] or [Drop
error] alarm is cancelled.
Press
or
to
cancel alarm.
[Please reset
time]
Low
RTC time is reset when the
pump is on.
Reset the time or press
to cancel alarm.
[Para.Unconfirm
ed]
Low
The alarm is triggered if no
operation is performed within
10s under the parameters
setting status.
Press any key to cancel
alarm.
[No Battery
Inserted]
Low
No battery inserted or battery
disengaged.
Correctly insert the battery.
[Pump serial
number conflict]
Low
After inserted to BeneFusion
DS5 Infusion Supervision
System, the pump has the
same serial number with other
pump. The alarm is triggered
by pumps with the conflicting
serial number.
1. Modify pump serial
number and ensure
different serial numbers in
BeneFusion DS5 Infusion
Supervision System.
2. Pull out the pumps with
the conflicting serial
number.
NOTE
All alarm sounds can be paused by pressing
, except for the
circumstance of [No Battery].