C-4
Alarm
Information
Alarm
Level
Reason
Countermeasure
over Wi-Fi, the network
communication is abnormally
interrupted for 3 minutes.
After the alarm is triggered,
infusion of the pump will not
be influenced, and the pump
continues infusion.
BeneFusion CS5 Infusion
Supervision System.
[KVO Running]
Low
The alarm is triggered after
[VTBI Complete] alarm is
triggered and in KVO mode.
1. Press
to stop
infusion and cancel alarm.
2. Or the [KVO Finish]
alarm is triggered when
reach the KVO infusion
time.
[Infusion
interrupted]
Low
The alarm is triggered after
any of the [Occlusion], [Air in
line], [Accumulate air], [Door
opened], [No Battery] or
[Drop error] alarm is
cancelled.
Press
or
to
cancel alarm.
[Please reset
time]
Low
RTC time is reset when the
pump is on.
Reset the time or press
to cancel alarm.
[Para.Unconfirm
ed]
Low
The alarm is triggered if no
operation is performed within
10s under the parameters
setting status.
Press any key to cancel
alarm.
NOTE
All alarm sounds can be paused by pressing
, except for the
circumstance of [No Battery].