11-2
11.2 Maintenance Plan
The following tasks must be conducted by professional maintenance personnel
approved by the company. Please contact the company if the following
maintenances are needed. Must clean and disinfect the device before the test or
maintenance.
Inspection/Maintenance Items
Frequency
Perform a safety inspection according
to the IEC60601-1 standard.
Once every two years. Perform after the
board is changed or the infusion pump is
accidentally dropped.
Preventive maintenance (refers to the
Maintenance Manual for pressure
calibration, sensor calibration, and
pump inspection).
Once every two years, or when you
suspect the occlusion alarm is abnormal,
the flow volume is inaccurate, or the
infusion set is incorrectly identified.
11.3 View Information
Select [
Main Menu
]→ [
History Record
]. In the [
History Record
] interface, you can
view the infusion parameters, alarm information, operation information and other
information.
Select [
Main Menu
]→[
System Options
]→[
Version Information
]. In the [
Version
Information
] interface, you can view the Software Version, Brand Library Version
and other versions.
11.4 Safe Disposal and Recycling
Please contact the company for related information about safe disposal and
recycling.