3-5
2. Connect one end of the power cord to the device's AC power supply port, and
check that internal and surrounding AC power supply port without liquid drug
and other residues.
3. Insert the other end of the power cord into the three-plug connector. Check that
the AC/DC indicator light is on.
WARNING
The earthing wire in the three-plug connector should be grounded, if
there is a doubt whether the AC power system is grounded or not,
please adopt the built-in battery and contact an electrical technician at
the hospital or the company.
Do not touch the power plug with wet or moist hands! If there is a liquid
drug or residue on or around the both ends of the power cord, power
socket or plug, device's AC power supply port, the user should
completely clean and dry the area before plugging into the power
supply, or accidents or injuries may result!
NOTE
Compatible power supply: 100
–240V
~
, 50/60Hz.
The AC power cable should be correctly inserted and secured into the
socket.
Removing power cord is to disconnect device from power supply.
Please ensure suitable clearance around the device to facilitate connect
and remove power cord.
3.1.6 Install and Operate the Drop Sensor (Optional)
NOTE
This section should be used with the optional drop sensor. The user may
skip the instructions in this section if a drop sensor is not included with
the infusion pump.