A-6
Mode, Sequential Mode
Pressure
4 Levels are adjustable, respectively
are:
150±125mmHg (20.0±16.7kPa),
300±125mmHg (40.0±16.7kPa),
525±125mmHg (70.0±16.7kPa),
900±180mmHg (120.0±24.0kPa).
Maximum occlusion pressure is about
1080mmHg.
Note 1: The detected pressure of the
infusion system (infusion pump and
infusion set) is affected by inner and
outer diameter, material, elasticity of
the infusion set and other factors.
Therefore, infusion sets of different
brand and model may differ in the
detected pressure range.
Note 2: The above declared pressure is
based on BOON A2 and B.Braun
Intrafix Safeset infusion sets,
temperature of 20±2
℃
.
525mmHg
Air detection
6 Levels are adjustable, respectively
are (20,
50, 100, 250, 500, 800) μl
Sensitivity of single air bubble is 20
μl.
Note 1: The detected air detection of
the infusion system (infusion pump and
infusion set) is affected by inner and
outer diameter, material, elasticity of
the infusion set and other factors.
Therefore, infusion sets of different
brand and model may differ in the
detected air detection.
Note 2: The above declared air
detection is based on BOON A2 and
B.Braun Intrafix Safeset infusion sets,
temperature of 20±2
℃
.
100
μl
Accumulate air
0.1-4ml
1.5ml
KVO Rate
0.1-5.0ml/h
0.5ml/h
Time Near End
Off, 1-30min
when the time is <10min, step for 1min,
and step for 5min when the time is
≥10min.
3min
Summary of Contents for BeneFusion VP3 Vet
Page 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Page 2: ......
Page 73: ...A 11 Unit of Rate ml h...
Page 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Page 91: ......
Page 92: ......