4-5
4.2.6 Parameter Memory Function
In clinical treatments, the medical staffs need to initiate the infusion as soon as
possible during emergency situations, infuse the liquid drug into the animal's body
in the shortest time possible, and set the detailed parameters later during infusion.
1. Select
[
Main Menu
]→[
User Maintenance
]→Input User Maintenance
Password→[
Para.
Memory
].
2. Select [
Para. Memory
]→[
On
]. If [
Off
] is selected, the following steps cannot
be conducted.
3. After selecting the infusion set brand, the previous infusion screen will appear,
the previous therapy parameters will be loaded, the users can use the
previous treatment parameters.
4.2.7 Select Infusion Mode
Press
to enter [
Main Menu
]
→[
Select Mode
]. On this interface, users can
press
and
to select infusion mode. Please refer to
Chapter 5
Infusion Mode
for detailed introduction of each infusion mode.
4.2.8 Purge Air
During infusion, the user should prevent air bubbles from entering the blood with
the liquid drug, which may form an aeroembolism and put animals in serious
danger. Therefore, air bubbles in the infusion tube should be eliminated before
the infusion. Under non-running status of any infusion mode, press
to enter
[
Purge Air
] prompt screen. Hold down
to enter [
Purge Air
] running screen,
release
after the air bubbles are purged.
WARNING
Please disconnect the pump from animal before purge. Otherwise, the
animal will be in serious danger
!