3-4
3.1.4 Fix BeneFusion DS3 Infusion Supervision System
(Optional)
The device can be located in the BeneFusion DS3 Infusion Supervision System.
NOTE
All components of the system are suitable for use in animal
environment.
Removing power cord is to disconnect equipment from power supply.
Please ensure suitable clearance around the System to facilitate connect
and remove power cord.
System assembly and refit during life period must be carried out by
maintenance technicians whom are trained and licensed by the
manufacturer, and evaluated according to the specified IEC60601-1.
Please contact the company if you have any queries.
Please ensure not simultaneously touch animal and device to prevent
animal leakage current from exceeding the requirements specified by
the standard.
Only devices designated by the manufacturer can be connected to the
system. Based on animal safety, do not insert devices that are not
specified by the manufacturer into the system.
3.1.5 Connect the AC Power Source
To connect the pump to the AC power source, follow this procedure:
1. Always use the accompanying power cord delivered with the pump, and check
that both ends of the power cord and surroundings without liquid drug and other
residues.