1-3
should be taken.
It is recommended that infusion pump is used with bulit-in infusion
sets . When use of non-bulit-in infusion sets, please make sure to
confirm relevant infusion performance (such as accuracy, air bubble and
pressure) on infusion pump, and contact the company for calibration
service, the infusion sets can only be used after confirmation, otherwise
Mindray Scientific is not responsible for infusion performance (such as
accuracy, air bubble and pressure) and relevant alarm function of the
infusion pump. Its accuracy cannot be guaranteed when the pump is
used with an infusion set without calibration.
When operating the pump and under non-running status, please
carefully check whether there is leakage from the tube export of the
infusion set, and drop needle of the drip chamber. If any leakage is
found, please contact the manufacturer for maintenance and
confirmation.
Do not touch the animal when connecting the peripheral equipment via
the input/output signal ports to prevent animal leakage current from
exceeding the requirements specified by the standard.
In the process of defibrillation, do not touch animal and other
non-defibrillation equipments to prevent electric shock damage, and
defibrillation will not affect the basic performance (such as infusion
accuracy, alarm and signal transmission) of the pump.
The equipment use a mains plug as isolation means to the mains power.
Do not locate the equipment in a place difficult to operate the mains
plug.