7-3
pump and CIMS depends on whether the network connection is successful,
operators are unable to observe the operation status of the pump in real time when
the communication is interrupted. After the network connection settings of the pump
and CIMS are modified, operators shall reset the network connection as required in
the manual to ensure the communication of the pump and CIMS are restored.
NOTE
Wireless security transmission distance is no more than 50 meters.
2.4 GHz Wi-Fi frequency range, and 802.11b/g/n wireless standard are
supported.
The settings of the wireless network must be conducted by technicians
approved by the company or maintenance staff designated by the
company.
It is recommended that not access the pump to public network. If the
pump is required to access to public network, firewall shall be added to
prevent cyber attacks.
7.5 Data Export
To export the data in the infusion pump, please refer to the following steps:
1. Log on PC tools, and connect the PC to the infusion pump;
2. When the infusion pump is in working communication with the PC, the PC
automatically reads all the data in the pump;
3. Select [
History Record
] in PC tools, and export data.