6-1
6
Setting Parameters
6.1 KVO
KVO (Keep Vein Open) means to keep the vein open, during which the infusion
pump continues infusion at a very low rate after finishing the infusion in order to
prevent blood backflow or vascular occlusion.
1. Select [
Main Menu
]→[
General Options
]→[
KVO Rate
].
2. Select [
KVO Rate
]: 0.1-5.0ml/h is adjustable.
6.2 Drug Library
The product is configured with a drug library with drugs available for users to select
from.
1. Select
[
Main Menu
]→[
User Maintenance
]→Input User Maintenance
Password→[
Drug Library
].
2. Select [
Drug library
]→[
On
], [
Off
] indicates switching off the function.
3. Select [
Drug
] on the parameters setting interface of Rate Mode. After the drug
is selected, its name will appear on the Run screen.
List of Drugs
Metoprolol
Dobutamine
Insulin
Flumazenil
Amiodarone
Dopamine
Sodium Nitroprusside
Fentanyl
Esmolol
Epinephrine
Diazoxide
Remifentanil
Lidocaine
Isoprenaline
Nimodipine
Morphine
Propafenone
Norepinephrine
Mannitol
Diazepam
Nitroglycerin
Atropin
Furosemide
Midazolam
Nicardipine
Succinylcholine
Phentolamine
Magnesium
sulfate
Isosorbide dinitrate
Pancuronium
Urapidil
Dexmedetomidine
Hydrochloride
Diltiazem
Aminophylin
Metaraminol Bitartrate
potassium
chloride
Digoxin
Heparin
Propofol
vasopressin
Summary of Contents for BeneFusion VP3 Vet
Page 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Page 2: ......
Page 73: ...A 11 Unit of Rate ml h...
Page 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Page 91: ......
Page 92: ......